Abstract
Purpose
To compare variations in surface morphology, as studied by scanning electron microscopy (SEM), of explanted intraocular lenses (IOLs) concerning the cause leading to the explantation surgery.
Methods
In this prospective multicenter study, explanted IOLs were analyzed by SEM and energy-dispersive X-ray spectroscopy. The IOLs were explanted in the centers of the research group from 2006 to 2012. The primary procedure was phacoemulsification in all cases.
Results
The study evaluated 40 IOLs. The main causes for explantation were IOL dislocation, refractive error, and IOL opacification. Those explanted due to dislocation demonstrated calcifications in 8 lenses (50%), salt precipitates in 6 cases (37.5%), and erythrocytes and fibrosis/fibroblasts in 2 cases (12.5%). In the refractive error cases, the SEM showed proteins in 5 cases (45.5%) and salt precipitates in 4 lenses (36.4%). In IOL opacification, the findings were calcifications in 2 of the 3 lenses (66.6%) and proteins in 2 lenses (66.6%).
Conclusions
A marked variation in surface changes was observed by SEM. Findings did not correlate with cause for explantation. Scanning electron microscopy is a useful tool that provides exclusive information regarding the IOL biotolerance and its interactions with surrounding tissues.