Abstract
BACKGROUND:
Intrapulmonary percussive ventilation (IPV) is frequently used for airway clearance, together with delivery of aerosolized medications. Drug delivery via IPV alone increases with decreasing percussion frequency and correlates with tidal volume (
METHODS:
An adult ventilator circuit was used for IPV–invasive ventilation. A normal or a diseased lung model was configured to airway resistance of 5 cm H2O/L/s and lung compliance of 100 mL/cm H2O or to airway resistance of 20 cm H2O/L/s and lung compliance of 50 mL/cm H2O, respectively. The ventilator settings were the following: pressure control continuous mandatory ventilation mode, 10 breaths/min; PEEP, 5 cm H2O;
RESULTS:
The maximum efficiency of albuterol delivery via IPV–invasive ventilation was not different from that via IPV alone (mean ± SD of loading dose, 3.7 ± 0.2% vs 4.2 ± 0.3%, respectively; P = .12). The mean ± SD albuterol delivery efficiency with IPV–invasive ventilation was lower for the diseased lung model versus the normal model (1.6 ± 0.3% vs 3.2 ± 0.5%; P < .001), which increased with decreasing percussion frequency. In contrast, the mean ± SD
CONCLUSIONS:
Albuterol delivery via IPV–invasive ventilation was modulated by percussion frequency but was not increased with increasing