BACKGROUND:
Peak flow testing is a common procedure performed in ambulatory care. There are currently no data regarding aerosol generation during this procedure. Given the ongoing debate regarding the potential for aerosol transmission of SARS-CoV-2, we aimed to quantify and characterize aerosol generation during peak flow testing.
METHODS:
Five healthy volunteers performed peak flow maneuvers in a particle-free laboratory space. Two devices continuously sampled the ambient air during the procedure. One device can detect ultrafine particles 0.02–1
RESULTS:
Ultrafine particles (0.02–1
CONCLUSIONS:
In this study, aerosol generation was present during peak flow testing, but concentrations were small compared to the background particle concentration in the ambient clinical environment. Surgical masks and eye protection are likely sufficient infection control measures during peak expiratory flow testing in asymptomatic patients with well controlled respiratory symptoms, but COVID-19 testing remains prudent in patients with acute respiratory symptoms prior to evaluation and peak expiratory flow assessment while the community prevalence of SARS-CoV-2 cases remains high.