Abstract
BACKGROUND:
Assisted coughing via mechanical in-exsufflation (MI-E) is a first-line treatment for secretion management in patients with amyotrophic lateral sclerosis (ALS) with unassisted CPF < 4.25 L/s. Some devices enable oscillations to be added to MI-E (MI-E+O). We sought to determine whether adding oscillations to MI-E enables a reduction in the use of invasive secretion management procedures (ie, bronchoscopy or tracheostomy) in subjects with ALS.
METHODS:
We conducted a 12-month, prospective, randomized follow-up study of subjects with ALS for whom assisted coughing techniques were indicated. One group was treated with oscillations in addition to MI-E (MI-E+O), and the other group was treated with conventional MI-E.
RESULTS:
29 subjects were included in the MI-E group and 27 subjects were included in the MI-E+O group. Five subjects (8.9%) required invasive techniques for secretion management (3 in the MI-E group and 2 in the MI-E+O group,
CONCLUSION:
Adding oscillations to MI-E did not enable a reduction in the need to perform invasive procedures for secretion management in subjects with ALS.