Abstract
BACKGROUND:
Widespread use of transcutaneous PCO2 (PtcCO2 ) monitoring is currently limited by concerns many practitioners have regarding accuracy. We compared the accuracy of PtcCO2 with that of PaCO2 measurements in critically ill children, and we investigated whether clinical conditions associated with low cardiac output or increased subcutaneous tissue affect this accuracy.
METHODS:
We performed a single-center prospective study of critically ill children placed on transcutaneous monitoring.
RESULTS:
There were 184 children enrolled with paired PaCO2
and PtcCO2
values. Subjects had a median age of 31.8 mo (interquartile range 3.5–123.3 mo). Most children were mechanically ventilated (
CONCLUSIONS:
PtcCO2 provides an acceptable estimate of PaCO2 in many critically ill children, including those with clinical conditions that may be associated with low cardiac output or increased subcutaneous tissue, although it does not perform as well in children with cyanotic heart disease. PtcCO2 may be a useful adjunct monitoring method, but it cannot reliably replace PaCO2 measurement.