Abstract
OBJECTIVE:
Evaluate the effects of continuous positive airway pressure (CPAP)/positive endexpiratory pressure (PEEP) and pressure support ventilation (PSV) on work of breathing (WOB).
METHODS:
With 13 anesthetized lambs we measured WOB with an esophageal balloon and flow signals. All the animals were sedated, intubated, and ventilated, using 2 pediatric ventilators (Servo 300 and VIP Bird). Ventilator settings were CPAP of 0, 5, and 10 cm H2O and PSV of 5 and 10 cm H2O with PEEP of 0, 5, and 10 cm H2O. Data were analyzed with 2-way analysis of variance.
RESULTS:
With the Servo 300 the total WOB (WOBT) increased between CPAP/PEEP of 0 and 10 cm H2O (p ≤ 0.0001) and between CPAP/PEEP of 5 and 10 cm H2O (p ≤ 0.0002). With the Servo 300 the addition of PSV decreased WOBT (p ≤ 0.003). With the VIP Bird the WOBT significantly increased between CPAP/PEEP of 0 and 10 cm H2O (p ≤ 0.02) and between CPAP/PEEP of 5 and 10 cm H2O (p ≤ 0.03). With PSV the WOBT was lower only at PSV 10 cm H2O (p ≤ 0.0001).
CONCLUSIONS:
WOBT increased as CPAP/PEEP was increased, and PSV lowered WOBT. We hypothesize that in a healthy animal model increased CPAP/PEEP may cause alveolar overdistention.