Hudson prongs (Hudson RCI, Teleflex Medical, Research Triangle Park, North Carolina) (a device to deliver nasal continuous positive airway pressure) are often secured with a “Velcro mustache” in neonatal intensive care units. We report 2 premature infants who required bag-and-mask ventilation while on Hudson prongs secured with a Velcro mustache. Effective ventilation was achieved only after removing the Velcro mustache.