Abstract
Background
Bioimpedance spectroscopy (BIS) is a valuable tool to assess nutrition and volume status in peritoneal dialysis (PD) patients. However, data about the influence of intraperitoneal fluid on body composition measures are conflicting, and there is no clear consensus about whether the abdomen should be drained before the procedure. We designed a comparison study to detect the influence of intra-abdominal fluid on BIS results.
Methods
We performed 73 pairs of BIS measurements in 34 stable PD patients, first with the peritoneum filled with a 1.36% glucose dialysate solution and then after the solution was drained. Patients stayed in the supine position for 10 minutes before the BIS procedure, and the electrodes were not moved between measures. Clinical and demographic data were collected, as were analytic parameters of nutrition and volume status.
Results
Fluid overload is overestimated when BIS is performed with a full abdomen (1.82 ± 1.73 L vs 1.64 ± 1.68 L,
Conclusions
Fluid overload is overestimated by BIS when performed in patients with dialysate in the peritoneum. The observed differences are greater in younger, more poorly nourished, or less overhydrated patients. If more precise results are required, we suggest that the abdomen be drained before BIS is performed.
Get full access to this article