Abstract
BACKGROUND:
Robotic assistance in total hip arthroplasty (RA-THA) has been shown to minimize laterality-based differences in acetabular cup positioning.
OBJECTIVE:
To determine if the use of a novel, fluoroscopy-based RA-THA system mitigates differences in acetabular cup placement between left (L) and right (R) side hip procedures, when compared to manual, fluoroscopic-assisted technique.
METHODS:
We conducted a retrospective review of 106 consecutive mTHA (40 L/66 R) and 102 RA-THA (48 L/54 R) primary direct anterior approach procedures. All cases were performed by a single right-hand-dominant surgeon, for a pre-operative diagnosis of osteoarthritis, avascular necrosis, or rheumatoid arthritis. Outcomes included acetabular cup inclination and anteversion, and the proportion of cups within the Lewinnek safe-zone.
RESULTS:
The average inclination of mTHA L cases was smaller than that of mTHA R cases (41.10∘
CONCLUSION:
Use of a novel, fluoroscopy-based robotic system mitigates laterality-based differences in acetabular cup placement that were observed in a manual, fluoroscopic-assisted cohort.