BACKGROUND:
Newborn hypothermia at birth remains as global challenge across all settings. The prevention of delivery room hypothermia at birth could potentially reduce neonatal morbidity and mortality.
OBJECTIVE:
To compare the heat conservation efficacy of Neohelp and Neowrap and evaluate the heat production efficacy of trans-warmer infant mattress (TWM) in a laboratory setting.
METHODS:
A beaker of water was heated at 60
RESULTS:
Neowrap took 863 seconds for the temperature to drop from 37∘C to 35∘C, compared with 941 seconds with Neohelp. When activated TWM reached a maximum temperature of 39.3
CONCLUSION:
Compared to Neowrap, Neohelp had better heat conservation properties. Activating the metal disk from the TWM center would deliver quicker heat.