Abstract
Computer Tomography (CT) guided percutaneous bone biopsy has been established as a minimally invasive method for the rapidly obtaining samples from osseous lesions for histological examination. With this method, an overall accuracy of 90% can be obtained without a significant radiation exposure. The aim of this study was to develop a navigated sleeve for the Jamshidi needle in order to perform navigated, percutaneous bone biopsies with the help of Iso-C-3D. Utilizing this navigated biopsy needle, the procedure was analyzed for several parameters including precision, total operating time, and level of radiation exposure. By using the Iso-C-3D navigation the total operating time was 11.7 min (8–14.3) and radiation exposure was 1.6 (0.67–2.12 min). This study demonstrates that navigated percutaneous bone biopsies can be obtained fast and with low radiation exposure. The feasibility of Iso-C-3D navigated bone biopsies is tempered by the requirement a rigidly fixed dynamic reference frame.
Get full access to this article
View all access options for this article.