Abstract
BACKGROUND:
Up to date there is no intramedullary, biodegradable osteosynthesis commercially available to treat non-comminuted midshaft fractures of small hollow bones applying not only a stable osteosynthesis but an additional axial compression to the fracture site.
OBJECTIVE AND METHODS:
Therefore we (1) designed different implant profiles and simulated the inner tension/volume using CAD. (2) Thereafter we manufactured a prototype with the best volume/tension-ratio using 70:30 poly-(L-lactide-co-D, L-lactide) (PLLA/PDLLA) and poly-
RESULTS:
The 3-point-bending-test-results showed that mean failure-force of PCL-tubes was 57.94
CONCLUSION:
The study showed that the new type of biodegradable, intramedullary tension-osteosynthesis made of PLLA/PDLLA is even more stable than common plate osteosynthesis in a small-hallow-bone-model. Further in vivo investigation should be performed to evaluate the surgical technique and long-term healing process of the bone and biodegradation process of the implant.