Abstract
BACKGROUND:
Implantable medical sensors for monitoring and transmitting physiological signals like blood glucose, blood oxygen, electrocardiogram, and endoscopic video present a new way for health care and disease prevention. Nevertheless, the signals transmitted by implantable sensors undergo significant attenuation as they propagate through various biological tissue layers.
OBJECTIVE:
This paper mainly aims to investigate the power loss of an out-to-in body wireless radio frequency link at 2.45 GHz.
METHODS:
Two simulation models including the single-layer human tissue model and three-layer human tissue model were established, applying the finite element method (FEM). Two experiments using physiological saline and excised porcine tissue were conducted to measure the power loss of a wireless radio frequency link at 2.45 GHz. Various communication distances and implantation depths were investigated in our study.
RESULTS:
The results from our measurements show that each 2 cm increase in implantation depth will result in an additional power loss of about 10 dB. The largest difference in values obtained from the measurements and the simulations is within 4 dB, which indicates that the experiments are in good agreement with the simulations.
CONCLUSIONS:
These results are significant for the estimate of how electromagnetic energy changes after propagating through human tissues, which can be used as a reference for the link budget of transceivers or other implantable medical devices.
Get full access to this article
View all access options for this article.