Abstract
INTRODUCTION:
This study compared the inter-instrument reliability of the Jamar Hand Dynamometer (Jamar) to the BTE EVJ Handgrip tool (EVJ). The EVJ is a new digital handgrip instrument designed by BTE Technologies Inc. (BTE Tech) utilizing Bluetooth technology to automatically upload data to a computer or electronic health record.
METHOD:
This repeated measure study engaged 338 participants (
RESULTS:
Comparative statistics included ICC values (0.81–0.84), Pearson R (correlation coefficient) (0.68–0.77), and Bland Altman plots (93–95% of data within 2 SD), indicating good inter-instrument agreement.
CONCLUSIONS:
The EVJ Handgrip, developed by BTE Technologies, demonstrated good inter-instrument reliability with the Jamar Hand Dynamometer and may be reliable to use when referencing the published normative values in the clinical environment.