Abstract
Objective: Oxygen saturation (SpO
Study design: 29 patients were monitored for 2 hours each with the study monitor. Upper and lower oxygen saturation alarm limits were determined individually and manually recorded and preset on the test oximeter for later analysis of alarm events. Test oximeter display values were not used for patient management. Audible alarms were silenced throughout the study.
Results: Alarm events were recorded while the SatSeconds™ alarm management feature was set at various points. With application of the SatSeconds™ alarm management feature, there was an overall decrease in alarms to 40% with a setting of 50 SatSeconds™.
Conclusion: Application of an integrated alarm system such as SatSeconds™ alarm management feature allows caregivers to respond to clinically relevant alarms.
Keywords
Get full access to this article
View all access options for this article.