Abstract
BACKGROUND:
Two-dimensional (2D) - shear wave elastography (SWE) has made promising advances in the diagnostic of breast lesions. However, few studies have assessed whether the diagnostic effectiveness of different platforms employing 2D-SWE is equal or different.
OBJECTIVE:
To compare the diagnostic effectiveness of 2D-SWE techniques from two different systems in differentiating malignant breast lesions from benign ones.
METHODS:
A total of 84 breast lesions were retrospectively analyzed by experienced radiologists using 2D-SWE on two ultrasound systems,
RESULTS:
Among the 84 lesions in this study, 66 (78.6%) were benign and 18 (21.4%) were malignant. E-max in system-1 showed the best diagnostic performance with a cut-off value of 174.5 kPa with the associated sensitivity and specificity of 100.0% and 80.3% respectively. Meanwhile, E-sd in system-2 displayed the best diagnostic performance with a cut-off value of 12.7 kPa, with the associated sensitivity and specificity of 94.4% and 80.3% respectively. The diagnostic performance of the two 2D-SWE systems was not statistically different according to receiver operating characteristic curve (ROC) analysis of E-max, E-mean, and E-sd.
CONCLUSION:
For identifying breast lesions, system-1 and system-2 appear to be similar in diagnostic performance. However, different cut-off values for different parameters might be selected to obtain the best diagnostic performance for the two 2D-SWE systems.