Abstract
PURPOSE:
To evaluate the use of contrast-enhanced ultrasound (CEUS) after irreversible electroporation (IRE) of prostatic cancer tissue to assess the ablation status by depicting microvascularisation in the ablation area.
MATERIALS AND METHODS:
Retrospective evaluation of CEUS of 13 patients (mean age: 61.4 ± 7.5 years) with histologically confirmed prostatic cancer who underwent percutaneous IRE. In the course of clinical routine, the tumor lesions were documented before, immediately after, and 1 day after the ablationusing color-coded transabdominal and transrectal CEUS. The obtained image data (DICOM loops and images) were subsequently evaluated by 2 experienced radiologists and assessed with regard to micro vascularisation by means of a 5-point scale.
RESULTS:
CEUS images showed significantly reduced microcirculation of the lesions (mean 0.9 ± 0.6 cm (0.5–1.5 cm) after IRE. Microcirculation was reduced from 2.15 ± 0.56 prior to ablation to 0.65 ± 0.63 (p < 0.001) immediately after the ablation and to 0.27 ± 0.44 one day after IRE (p < 0.001).
CONCLUSION:
This study showed rapid and significant reduction of the microcirculation in the ablation area afterpercutaneous IRE of prostatic cancer tissue.
Keywords
Get full access to this article
View all access options for this article.
