Abstract
In connection with a problem dealing with the effects of various anesthetic gases on small laboratory animals (rabbits), it was necessary in the interest of safety and economy to construct an apparatus which could be attached to any standard clinical gas machine. After considerable difficulty, an apparatus meeting these requirements was designed. This apparatus makes it possible to supply anesthetic gases to the animal in rather constant concentrations.
The apparatus (Fig. 1) consists of a small soda-lime cannister and 2 flutter valves, mounted on a base 8 by 6 inches. Accurate quantities of the gases are measured by a large gas machine and are delivered into the small machine through the inlet tubing (A). The gases then pass over the flutter valve (B) and are delivered into the face mask. The exhaled gases pass through the flutter valve (F) and then through a small soda-lime cannister into the rebreathing bag. On inspiration, the mixture of gases from the rebreathing bag passes through the inspiratory flutter valve (B) to the animal. The small chamber above this flutter valve acts as a mixing chamber for the fresh and the rebreathed gases.
This constitutes a closed carbon dioxide absorption circuit and allows rebreathing of the anesthetic gases. The flutter valves (B and F) alternately open and close on inhalation and exhalation, thus compelling all gases to pass in one direction only. The rebreathing bag is a small rubber bag of 400 cc capacity. The metal bell-shaped face mask is covered with a rubber diaphragm in the center of which is a small opening to admit the nose of the animal. The tubing has an inside diameter of 8 mm. This apparatus permits small animals to be anesthetized with a minimal amount of anesthetic gases.
Get full access to this article
View all access options for this article.