Abstract
In an attempt to find a method of securing accurate time curves of red blood corpuscle hemolysis by various hemolytic agents, an apparatus was devised in which a constant beam of light was passed through a dilute blood suspension contained in a revolving glass chamber. The emergent beam was allowed to fall on a potassium hydride photo-electric cell, and the current so set up in the cell was suitably measured. 1 Using, under standard conditions, supposedly unchanging blood suspensions in various isotonic fluids including saline, Brinkman's solution, 2 several modifications of it, a special phosphate solution, and a modified Locke's solution, 3 marked spontaneous variations in the emergent beam were noted during the first one to three hours after the preparation of the suspensions. Usually these variations consisted of a gradual progressive diminution in light intensity; sometimes an increase occurred.
Studies of the pH of the diluting fluids before and after the experiments revealed no relationship to the light transmissibility of the suspensions. Microscopic observations of the cell shape on flamed slides 4 , 5 showed various types of crenation but no correlation between changes in cell shape and in light transmission. The cell volumes of the suspensions were determined in a special hematocrit, and in general light transmission varied inversely as total cell volume, but no definite quantitative relation could be determined.
The following technique was eventually developed to obtain a constant red cell suspension. Washed cells to represent 0.5 per cent whole blood were added sterilely to Brinkman's solution containing 0.1 per cent NaHCO3, and allowed to stand 18-24 hours in paraffined flasks. All of 54 such suspensions showed less than 2 per cent change in light transmission during a three hour period.
Get full access to this article
View all access options for this article.