Abstract
Purpose:
To confirm the performance and safety of the 25-cm Viabahn endoprosthesis with Propaten bioactive surface when used in the treatment of de novo and/or restenotic TransAtlantic Inter-Society Consensus II C and D lesions of the superficial femoral artery and proximal popliteal artery.
Methods:
The 25-cm Gore Viabahn Endoprosthesis study (
Results:
The median follow-up was 12.3 months (mean 12.3±0.6). Nine (12.7%) patients discontinued the study due to different reasons including 2 bypass surgeries. Angiography was available in 60 patients to determine the primary performance outcome, which was met in all cases. Two (2.8%) patients experienced a procedure-/device-related adverse event (dissection) during the 30-day follow-up. Kaplan-Meier estimates for 1-year primary and secondary patency were 67.0% [95% confidence interval (CI) 53.5% to 77.3%] and 96.9% (95% CI 88.0% to 99.2%), respectively. Changes in ankle-brachial index and Rutherford category at 1 and 12 months each showed sustained improvement.
Conclusion:
This study confirms that the 25-cm Viabahn endoprosthesis acutely performs as intended and is safe when used as indicated in complex femoropopliteal lesions. One-year primary and secondary patency rates are satisfying and comparable to historical prosthetic bypass graft outcomes.
Keywords
Get full access to this article