Abstract
Purpose
To evaluate the safety and efficacy of a balloon-mounted drug-eluting stent (DES) for recurrent carotid in-stent stenosis (ISS).
Methods
As part of our targeted carotid artery stenting (TARGET-CAS) protocol, neurological and ultrasound evaluations have been performed at 3, 6, and 12 months and then annually since 2001 in all carotid stent patients. For angiographically-confirmed >70% ISS, balloon angioplasty was performed as a first-line treatment. Recurrent ISS was treated with a 4.0-mm zotarolimus-eluting coronary stent (ZES) that was postdilated according to intravascular ultrasound imaging. Among the 1350 neuroprotected CAS procedures performed between January 2001 and March 2011, there were 7 (0.52%) patients (5 men; ages 51–72 years), all neurologically asymptomatic, with >70% recurrent ISS that occurred at 5 to 11 months after the initial balloon angioplasty treatment for ISS.
Results
ZES implantation under distal embolic protection was technically successful and uncomplicated. Angiographic stenosis was reduced from 84.6%±7.5% to 10.7%±3.6% (p <0.01). In 5 patients with ZES implanted fully within the self-expanding carotid stent, duplex ultrasound follow-up (mean 17 months, range 6–36) revealed no evidence of restenosis or stent fracture/deformation. In the 2 other patients, the ZES had been implanted for distal edge ISS such that the ZES protruded beyond the original carotid stent. This protruding segment of the ZES demonstrated deformation/kinking in both; in one, this led to symptomatic stent occlusion.
Conclusion
The use of coronary ZES in the treatment of recurrent carotid ISS is feasible and appears effective provided the ZES is placed entirely within the original stent. Placement of a coronary ZES outside the carotid stent scaffold should be avoided.
Keywords
Get full access to this article
View all access options for this article.