Abstract
Purpose
To report the 6-month results of a prospective randomized trial investigating angioplasty with paclitaxel-coated balloons (PCB) vs. plain balloon angioplasty (BA) for the treatment of failing native arteriovenous fistulae (AVF) or prosthetic arteriovenous grafts (AVG).
Methods
The enrollment criteria for this non-inferiority hypothesis trial included clinical signs of failing dialysis access with angiographic documentation of a significant venous stenotic lesion in patients with AVF or AVG circuits. From March to December 2010, 40 patients (29 men; mean age 64.1±14.3 years) were randomized to undergo either PCB dilation (n=20) or standard BA (n=20) of a stenosed venous outflow lesion. Regular angiographic follow-up was scheduled bimonthly. Study outcome measures included device success (<30% residual stenosis without postdilation), procedural success (<30% residual stenosis), and primary patency of the treated lesion (<50% angiographic restenosis and no need for any interim repeat procedures).
Results
Baseline and procedural variables were comparably distributed between both groups. Device success was 9/20 (45%) for the PCB device vs. 20/20 (100%) for standard control BA (p<0.001). Procedural success was 100% in both groups after further high-pressure post-dilation as necessary. There were no major or minor complications in either group. At 6 months, cumulative target lesion primary patency was significantly higher after PCB application (70% in PCB group vs. 25% in BA group, p<0.001; HR 0.30, 95% CI 0.12 to 0.71, p<0.006).
Conclusion
PCB angioplasty improves patency after angioplasty of venous stenoses of failing vascular access used for dialysis.
Keywords
Get full access to this article
View all access options for this article.