Abstract
Purpose
To quantitatively evaluate the impact of carotid stent cell design on vessel scaffolding by using patient-specific finite element analysis of carotid artery stenting (CAS).
Methods
The study was organized in 2 parts: (1) validation of a patient-specific finite element analysis of CAS and (2) evaluation of vessel scaffolding. Micro-computed tomography (CT) images of an open-cell stent deployed in a patient-specific silicone mock artery were compared with the corresponding finite element analysis results. This simulation was repeated for the closed-cell counterpart. In the second part, the stent strut distribution, as reflected by the inter-strut angles, was evaluated for both cell types in different vessel cross sections as a measure of scaffolding.
Results
The results of the patient-specific finite element analysis of CAS matched well with experimental stent deployment both qualitatively and quantitatively, demonstrating the reliability of the numerical approach. The measured inter-strut angles suggested that the closed-cell design provided superior vessel scaffolding compared to the open-cell counterpart. However, the full strut interconnection of the closed-cell design reduced the stent's ability to accommodate to the irregular eccentric profile of the vessel cross section, leading to a gap between the stent surface and the vessel wall.
Conclusion
Even though this study was limited to a single stent design and one vascular anatomy, the study confirmed the capability of dedicated computer simulations to predict differences in scaffolding by open- and closed-cell carotid artery stents. These simulations have the potential to be used in the design of novel carotid stents or for procedure planning.
Get full access to this article
View all access options for this article.