Abstract
A major concern in laboratories performing trace element determinations is sample contamination. Sensitivity can be enhanced for trace elements by using larger samples, but this approach may increase the risk of contamination due to the need for more involved preparation techniques. With the isolation of the sample from the laboratory environment, the potential for contamination is greatly reduced during sample decomposition. In this investigation, an environmental evaporation chamber (EEC) placed in a microwave digestion system was used for predigestion of biological and botanical samples prior to closed-vessel digestion. The EEC was also used for the evaporation of residual acid after completion of the digestion procedure. Data are presented for the analysis of standard reference materials along with detection limits and the results of a cross-contamination study for the described method. Index Headings: Inductively coupled plasma optical emission spectrometry; Biological samples; Botanical samples; Environmental evaporation chamber; Microwave digestion.