In this work, we describe the interfacing of a high-resolution chromatographic system featuring narrow bore gas chromatographic columns to a Fourier transform infrared spectrometer. Excellent chromatographic resolution, low nanogram sensitivity, and compound identification are demonstrated. Applications to the analysis of a petroleum distillation fraction and to a polymer sample, analyzed after pyrolysis, are presented. A vapor-phase library search system is used to identify compounds in both samples.
WallD. L. and MantzA. W., Appl. Spectrosc.31, 552 (1977).
15.
CoffeyP.MattsonD. R., and WrightJ. C., Am. Lab., 126 (May 1978).
16.
ShaferK. H.LucasS. V., and JakobsenR. J., J. Chromatogr. Sci.17, 464 (1979).
17.
CrawfordR. W.HirschfeldT.SanbornR. H., and WongC. M., Anal. Chem.54, 817 (1982).
18.
G. E. Adams, IBM Instruments, Inc., presented at the Exposition of the 1981 Pittsburgh Conference (unpublished).
19.
MolG. J.GritterR. J., and AdamsG. E., “Mass spectrometry of thermally treated polymers,” in Applications of Polymer Spectroscopy, BrameE. G.Jr., Ed. (Academic Press, New York, 1978), p. 257.