Iron analyses performed on used gas turbine lubricating oil samples by several variations of rotating disk–spark emission spectrography, atomic absorptiometry, and x-ray fluorescence spectrometry are compared with a quantitative wet chemical technique. The results indicate that emission spectrography with a cobalt internal standard, atomic absorptiometry with a nitrous oxide–acetylene flame, and x-ray fluorescence, both dispersive and nondispersive, are the most reliable instrumental methods. The iron contained in the samples exists primarily as particulates 1 µ or less in diameter.
JacksonD. R.SalamaC., and DunnR., Can. Spectrose.15, 17 (1970).
10.
GunnE. L., Advan. X-Ray Anal.11, 164 (1968).
11.
HaleC. C. and KingW. H.Jr., Anal. Chem.33, 74 (1961).
12.
GunnE. L., Anal. Chem.36, 2086 (1964).
13.
PackerL. L.WoodyB. A.BrutonW. A., and GehrkeR. J., “Ppm Measurements of Iron-in-Oil by Radioisotope X-ray Fluorescence,” Symp. Laser Energy X and Gamma Sources and Applications 3rd, Boston, Mass., June, 1970.
Proposed Spectrochemical Method for Wear and Contaminant Element in Used Diesel Lubrication Oils using a Rotating-Disk Electrode Technique (American Society for Testing Materials, Philadelphia, Pa., 1968), Appendix V: D-2-1968.
16.
O'ConnellR. F. (private communication).
17.
KriegeO. H. and WelcherG. G., “Major Component Analysis in Metallurgical Samples by Atomic Absorption Spectrometry,” Annu. Soc. Appl. Spectrosc. Mtg. 9th, New Orleans, October, 1970.
18.
SpragueS. and SlavinW., At. Absorption Newslett. 4, 293 (1965).
19.
SlavinW. and SlavinS., Appl. Spectrosc.23, 421 (1969).