The development of a direct-reading spectrometric method for the determination of Al2O3, CaO, Fe2O3, T1O2, MgO and Mn2O3 in high silica (> 92 per cent S1O2) is described. Once the effect of silica on the spectral intensities has been evaluated it has proved possible to ‘set up’ the instrument on silica-free standards without loss of accuracy. The results for Al2O3, T1O2, MgO and Mn2O3 are at least as accurate as existing chemical procedures; the results for CaO and Fe2O3 are of routine control level
Get full access to this article
View all access options for this article.
References
1.
TingleW. H. and MatochaC. K., Anal. Chem.30, 494, 1958.