The Liquid Metal Fuel Reactor Program requires rapid, accurate analyses of large numbers of bismuth metal samples. A spectrochemical solution procedure using a rotating disc technique with a Baird-Atomic Research Direct Reading Spectrograph is described. Yttrium is added as an internal standard. Magnesium and zirconium are determined with a precision of 2% of the amount present; the precision for iron, chromium, manganese, nickel and molybdenum is within 4–12%.
Get full access to this article
View all access options for this article.
References
1.
FeldmanC., ANAL. CHEM.21, 1041 (1949)
2.
DelaneyJ. C. and OwenL. E., ANAL CHEM.23, 577 (1951)
3.
HuntD. J. and PishG., U. S. Atomic Energy Commission, MLM- 801 (1953)
4.
SmartR. C. and WebbM. S. W., United Kingdom Atomic Energy Authority, AERE C/R 2117 (1958)
5.
CotterillJ. C., United Kingdom Atomic Energy Authority, AERE C/R 2456 (1958)
6.
OwensM. J. and WebbM. S. W., United Kingdom Atomic Energy Authority, AERE C/R 2115 (1957)