Abstract
A wireless hand-held Webpad device was used to review a sample set of cranial computerized tomography (CT) studies to assess its diagnostic capabilities and its feasibility as a portable diagnostic workstation for radiology. The data-set consisted of 30 head CT studies of emergency cases. Two neuroradiologists and a senior radiologist participated in the evaluation of the portable workstation. They used a Web-based viewer that we developed, which provided all the major functionalities required for radiological image review. The reported radiological findings and diagnoses were compared with a gold standard, comprising a set of diagnoses previously formulated by a consensus panel of radiologists who had reviewed the original studies. The diagnoses made using the Webpad were correct (no major discrepancies) in 82 out of 90 interpretations (91%), which is comparable to the accuracy reported in image review with a conventional radiological workstation. The average total working time per diagnosis was 5 min 25 s (range 2–12 min). The simplicity of use of the system and its low cost make it suitable for distributing radiological studies within hospital facilities.