Abstract
9.4 Tesla Fourier transform ion cyclotron resonance (FT-ICR) mass spectrometers (Bruker BioAPEX-94e) have been installed at the Division of Ion Physics, Uppsala University, and at the Department of Chemistry, University of Warwick. The BioAPEX-94e FT-ICR instrument is built around a high-field, superconducting magnet and a platform with easily interchangeable ion sources [matrix-assisted laser desorption/ionisation (MALDI), secondary ion mass spectrometry (SIMS), electrospray ionisation (ESI) and electron impact/chemical ionisation (EI/CI)]. In this paper a technical description of the instrument is given. Outstanding performance characteristics are demonstrated, notably clear resolution of C59N+ and C5813C2+ (mass difference 3.65 mDa) and mass measurement accuracy at the low ppm level. A wide range of applications in Warwick and Uppsala is described, demonstrating the versatility and high performance of the instrument.
Keywords
Get full access to this article
View all access options for this article.
