Abstract
Different phenylenediamines were used to explore anodic oxidation in solution during electrospray ionization (ESI) mass spectrometry analysis. In our experiments, a series of unknown ionic species was detected in the phenylenediamine solutions. Our results propose that reactions of phenylenediamines with species formed by anodic oxidation of typical ESI solvents during the electrospray ionization process such as formaldehyde are producing these peaks. Identification of these compounds inter alia suggests formal alkylation, a reaction not reported so far as a result of electrolytic oxidation in the prospective organic solvents.
Keywords
Get full access to this article
View all access options for this article.
