Brittle lacquer techniques and electric resistance strain gauge rosettes were used to measure strains on the stems of total hip replacements while loaded according to three standards. The stresses in these locations were also calculated and good agreement was found between the measured and calculated values.
The Sines equation was used to calculate the endurance limit stress. Again good agreement was found with observed results. Suggestions are made regarding the endurance testing of hip prostheses.
Get full access to this article
View all access options for this article.
References
1.
EvansF. G.LissnerH. R.PedersenH. E.Deformation studies of the femur under dynamic vertical loading. Anat. Rec., 1948, 101, 225–241.
2.
KalenR.Strains and stresses in the upper femur studied by the stresscoat method. Acta Orthop. Scand., 1961, 31, 103–113.
3.
MarkolfK. L.AmstutzH. C.A comparative experimental study of stresses in femoral total hip replacement components: The effects of prosthesis orientation and acrylic fixation. J. Biomech., 1976, 9, 73–79.
4.
ReubenJ. D.EismontF. J.BursteinA. H.WrightT. M.Comparative mechanical properties of forty-five total hip stems. Clin. Orthop. Rel. Res., 1979, 141, 55–65.
5.
SechlerE.Elasticity in engineering, 1952, Ch. 11, pp. 258–269. (Wiley, London).
6.
SinesG.OhgiG.Fatigue criteria under combined stresses or strains. Trans. ASME, J. Engng Matl Technol., 1981, 103, 82–90.
LeeC.LingR.Problems associated with fatigue testing. In Evaluation of artificial joints, 1977, Ch. 8, pp. 91–96 (Biological Engineering Society, London).
9.
BrownT. R. M.NicolA. C.PaulJ. P.Comparison of loads transmitted by Charnley and CAD Muller total hip arthroplasties. IMechE conference on Engineering and clinical aspects of endoprosthesis fixation, 1984, paper C210/84, pp. 63–68 (Mechanical Engineering Publications, London).
10.
MinerM. A.Cumulative damage in fatigue. J. Appl. Mech., 1945, 12, A159–A164.
11.
PaulJ. P.Forces transmitted by joints in the human body. Proc. Instn Mech. Engrs., 1966, 181, 8–15.
12.
WroblewskiB. M.The mechanism of fracture of the femoral prosthesis. Int. Orthop., 1979, 3, 137–139.
13.
CharnleyJ.Rotary forces on a femoral prosthesis. In Low friction arthroplasty of the hip—theory and practice, 1979, Ch. 9, pp. 121–122 (Springer Verlag, Berlin).