Abstract
Abstract
Ultra-high molecular weight polyethylene (UHMWPE) acetabular cups were tested against alumina-ceramic femoral heads using a new type of hip joint simulator according to ISO/FDIS 14242-1. Bovine serum as well as newborn calf serum were used as test fluids. Total polyethylene wear was determined by weight loss of the cups. In addition, wear depth and its distribution were recorded by means of a coordinate measurement system. Wear particle analysis and inspection of the worn polyethylene surfaces using light and scanning electron microscopy (SEM) were performed to analyse damage and identify the acting wear mechanisms.
The total wear rate was determined to be 22.07 ± 1.75 mg/106 cycles for the bovine serum group and 26.57 ± 3.55 mg/106 cycles for the calf serum group. Unexpectedly, the formation of two wear vectors corresponding to recent clinical findings was detected. Retrieved polyethylene wear debris was comparable in size and shape with clinical findings. The test method described by ISO/FDIS 14242-1 produced reliable and reproducible wear data using UHMWPE acetabular cups articulating against alumina-ceramic heads. In the authors' opinion, the lubricant composition should be described in more detail, since the protein and additive content seem to have a high impact on the wear results. It needs to be emphasized that the findings of this study cannot be regarded as a general validation of hip wear tests according to ISO/FDIS 14242-1 but are limited to the material combinations investigated herein. Further testing of other clinically relevant materials and interlaboratory ring tests must follow.
Get full access to this article
View all access options for this article.
