Abstract
Background: Reduction of atheroembolic complications during cardiopulmonary bypass remains a major challenge in cardiac surgery. New cannula tip designs may help to attenuate this problem by improved hydrodynamics.
Methods: Pressure gradients and back pressures of a new aortic cannula tip design were measured and compared with the Medos X-Flow, Sarns Soft-Flow and Argyle THI cannulae at various flow rates in a mock circulation followed by flow visualization.
Results: Pressure gradients were the lowest for the new cannula. Back pressures of the new cannula were up to 84% lower than for the Argyle cannula. The back pressure profile and flow visualization of the new cannula showed broad centric flow dispersion with a transcannula increase of flow area from 38 mm2 to 139 mm2.
Conclusions: The new design of an aortic cannula tip provides improved hydrodynamics, with low pressure gradients, low back pressures and a uniform central dispersion of flow, reducing the sandblasting effect.