Abstract
Purpose:
The purpose of this study is to determine the sensitivity and specificity of detecting age-related macular degeneration (AMD) using portable optical coherence tomography (OCT) operated by nonexpert photographers on undilated patients.
Methods:
In this case-control study, 92 individuals were recruited from the glaucoma and retina clinics at the Wilmer Eye Institute (Johns Hopkins University, Baltimore, MD). Using the portable iVue (Optovue, Inc, Fremont, CA) spectral-domain OCT (SD-OCT), 2 nonexpert photographers acquired retina map scans on undilated eyes of all participants. In total, 33 AMD eyes and 105 control eyes were evaluated and graded by ophthalmologists masked to the diagnoses.
Results:
Detection of AMD on the portable OCT by ophthalmologists exhibited sensitivities of 0.91 and 0.94 and specificities of 0.88 and 0.89, for graders 1 and 2, respectively. A strong intergrader agreement was observed (κ = 0.87).
Conclusions:
Nonexpert photographers can use a portable OCT on undilated eyes to acquire images for the detection of AMD. These findings present the potential utility of implementing a portable OCT in community screenings for earlier detection and treatment of disease.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.