Abstract
We evaluated Computed Tomography Angiography (CTA) of four different self-expandable stents in phantom model. Three stents ware made of nitinol and one was made of cobalt-based alloy. Each stent placed in plastic injector, filled with diluted iodinated contrast material, and placed on a plastic case filled with saline. CTA was performed on a four-detector CT scanner (detector collimation, 0.5mm: helical pitch 3.5). Axial and longitudinal reconstructions of CT imaging were compared with the in-stent attenuation. Three nitinol stents showed few artifact, and cobalt-based alloy stent was appeared artificial narrowing. In conclusion, CTA is useful for follow-up examination after carotid artery stenting (CAS) as a less invasive evaluation method.