Abstract
Degree of hearing loss in the infant is related to high-intensity or high-frequency sound (noise). We measured sound levels of 6 models of pneumatic nebulizers and 8 models of humidifiers at various flowrates, F102 settings, and water levels, when they were attached to incubators and oxygen hoods; we also evaluated sound levels from 2 models of ultrasonic nebulizers at various output settings. Among the pneumatic nebulizers, the recommended 58-dbA maximum sound level was exceeded by the Aquapak 621, Ohio Deluxe, and Puritan All-Purpose devices. The Bard Parker, Foregger, and Inspiron pneumatic nebulizers' sound levels were 58 dbA or below except during the dry water-level condition. Among the humidifiers, all produced sound levels below the recommended 58-dbA maximum except the Travenol humidifier; the Bennett Cascade, Conchapak, Foregger, and Hudson humidifiers produced mean sound levels in the 43.0- to 43.5-dbA range. Sound levels increased under all conditions when the devices were used with oxygen hoods. The ultrasonic nebulizers produced low sound levels, but the high-frequency ultrasound they also produce may be undesirable for infants; this question requires further investigation. (Beckham RW, Mishoe SC. Sound Levels inside Incubators and Oxygen Hoods Used with Nebulizers and Humidifiers.
Keywords
Get full access to this article
View all access options for this article.
