We evaluated the Waters MRM-6000 and Biergy VVR closed-circuit indirect calorimeters for accuracy and reproducibility of results in measuring oxygen uptake (
) and for performance during mechanical ventilation (MV). We also tested the MRM-6000 for accuracy and reproducibility in measuring carbon dioxide production (
). METHODS: We bench tested the devices to compare their measurements with those of a model during mechanical ventilation (MV). The model included a test lung, reference spirometer, vacuum pump, flow regulator, and adult volume ventilator. Tests were conducted with tidal volumes (VT) of 800-1600 mL at rates of 12-30 breaths/min; test-lung compliances were 0.01, 0.03, and 0.05 L/cm H2O [0.102, 0.306, and 0.510 L/kPa]; simulated
values were 109-844 mL/min. We also conducted clinical studies on 11 postoperative patients. RESULTS: No significant differences were found between bench-model and calorimeter
values for both calorimeters or between bench-model and MRM-6000
values. Good reproducibility of results was shown by a coefficient of variation of 1.5% for the MRM-6000 and 1.7% for the VVR. There were significant differences in displayed VT compared to VT delivered to the test lung, and individual adjustments had to be made to maintain the pretest level of mechanical ventilator support. CONCLUSIONS: The MRM-6000 and VVR calorimeters provide accurate and reproducible information during MV and perform adequately if the operator makes suitable adjustments during the test period.