In oxygen therapy, the behavior and effectiveness of high airflow oxygen-entrainment systems depends to a great extent on the peak inspiratory flowrate (PIFR) of the patient. The formula reported here, PIFR = πVE(TI + TE)/2TI, provides a more accurate calculation of peak inspiratory flow than does the often-used clinical rule of thumb, PIFR = 3. VE, and may be used to predict inspiratory fresh-gas flow requirements for ventilators and nonrebreathing circuits. (Respir Care 1985;30:766-770.)
Get full access to this article
View all access options for this article.
References
1.
MushinWW, Rendell-BakerL, ThompsonPW, MaplesonWW. Automatic ventilation of the lungs. 2nd ed. Oxford and Edinburgh: Blackwell, 1969:13.
2.
NunnJF. Applied respiratory physiology with special reference to anaesthesia. London: Butterworth, 1969.
3.
ScurrC, FeldmanS.Scientific foundations of anaesthesia. London: Heinemann, 1970:202.
4.
HillDW. Physics applied to anaesthesia. London: Butterworth, 1976:101.
5.
MaplesonWW. The elimination of rebreathing in various semi-closed anaesthetic systems. Br J Anaesth1954;26:323.
6.
PerchSA, RealeyAM. Effectiveness of the Servo SH 150 "artificial nose" humidifier: A case report. Respir Care1984;29:1009–1012.
7.
DemersRR. Oxygen delivery systems for use in acute respiratory failure. Respir Care1983;28:553–556.
8.
MonastRL, KayeW.Problems in delivering desired oxygen concentrations from jet nebulizers to patients via face tents. Respir Care1984;29:994–1000.
9.
ConnorsAF, McCaffreeDR, GrayBA. Effect of inspiratory flow rate on gas exchange during mechanical ventilation. Am Rev Respir Dis1981;124:537.
10.
GibsonRL, ComerPB, BeckhamRW, McGrawCP. Actual tracheal oxygen concentrations with commonly used oxygen equipment. Anesthesiology1976;44:71.