Abstract
Background:
This study evaluated performance and product labeling of CONTOUR® USB, a new blood glucose monitoring system (BGMS) with integrated diabetes management software and a universal serial bus (USB) port, in the hands of untrained lay users and health care professionals (HCPs).
Method:
Subjects and HCPs tested subject's finger stick capillary blood in parallel using CONTOUR USB meters; deep finger stick blood was tested on a Yellow Springs Instruments (YSI) glucose analyzer for reference. Duplicate results by both subjects and HCPs were obtained to assess system precision. System accuracy was assessed according to International Organization for Standardization (ISO) 15197:2003 guidelines [within ±15 mg/dl of mean YSI results (samples <75 mg/dl) and ±20% (samples ≥75 mg/dl)]. Clinical accuracy was determined by Parkes error grid analysis. Subject labeling comprehension was assessed by HCP ratings of subject proficiency. Key system features and ease-of-use were evaluated by subject questionnaires.
Results:
All subjects who completed the study (
Conclusions:
CONTOUR USB exceeded ISO 15197:2003 system performance criteria in the hands of untrained lay users. Subjects understood the product labeling, found the system easy to use, and successfully performed blood glucose testing.