Abstract
Background:
To prevent complications in diabetes, the proper management of blood glucose levels is essential. Since conventional glucose meters require pricking fingers or other areas of the skin, a noninvasive method for monitoring blood glucose levels is desired. Using a lightweight cymbal transducer array, this study was conducted to noninvasively determine the glucose levels of pigs having a similar size to humans.
Method:
Result:
In comparison, glucose levels determined by a cymbal array and biosensor system were close to those measured by a glucose meter. After a 20-minute ultrasound exposure at Isptp = 100 mW/cm2, the average glucose level determined by the ultrasound system was 175 ± 7 mg/dl, which is close to 166 ± 5 mg/dl measured by the glucose meter.
Conclusion:
Results indicate the feasibility of using a cymbal array for noninvasive glucose sensing on pigs having a similar size to humans. Further studies on the ultrasound conditions, such as frequency, intensity, and exposure time, will be continued for effective glucose sensing.