Abstract
Background
The aim of this study was to compare the reliability and agreement between two devices – Wii Fit Balance Board (WBB) versus Hand-Held Dynamometer (HHD) to measure isometric strength during the athletic shoulder (ASH) test in healthy amateur rugby players.
Methods
Fifteen males (23.73 ± 2.8 years) completed two testing sessions. Maximal isometric contractions using the dominant arm (D) and non-dominant arm (ND) against a WBB and HHD were assessed at three angles of abduction (180°, ‘I’; 135°, ‘Y’ and 90°, ‘T’), in a prone lying position.
Results
The results indicate a very large correlation between the HHD and the WBB. WBB provides acceptable reliability at I-Test D (CV = 9.97%, ICC = 0.88) and HHD in the I-Test D (CV = 8.90%, ICC = 0.94), I-Test ND (CV = 8.60%, ICC = 0.95) in peak strength values. The HHD is most reliable in D ASH I-Y-T (CV = 10.94%) and WBB (CV = 11.05%). In the ND ASH I-Y-T test, the HHD is the most reliable (CV = 12.5%) compared to the WBB (CV = 14.43%).
Conclusions
These results suggest that WBB is a reliable device to assess strength in the ASH test with a very large correlation with the HHD. WBB and HHD are two affordable devices to assess isometric shoulder strength.
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.