Abstract
Metal-on-metal articulations have fallen out of favour in larger joint replacements, but are still used in smaller joints. Coating the articulation has been suggested as one way of reducing wear. We compared a standard 6 mm CrCoMo articulation designed for the carpometacarpal joint of the thumb with a chromium nitride-coated version after 512,000 cycles in a joint simulator. A total of 6 articulations in each group were tested with a unidirectional load of 5 kg in Ringer’s solution. We found a statistically significant reduction in weight loss, amount of metallic wear produced and volumetric wear for the chromium nitride-coated articulation. Our findings support the use of chromium nitride coating in order to minimize the amount of metallic wear produced.