Coil embolization was performed in 86 cerebral aneurysm patients using two types of detachable platinum coils, IDC (Interlocking detachable coil) and GDC (Guglielmi electrical detachable coil). Results of IDC and GDC were compared. The occlusion rate of the aneurysm sac was similar. Coil compaction occurred frequent and early in cases with GDC. As clinical outcome, 94% of the patients in both group obtained good results. No bleeding or rebleeding occurred in the follow-up with IDC or GDC.
GuglielmiGViñuelaF: Electrothrombosis of succular aneurysms via endovascular approach. Part 2: Preliminary clinical experience. J Neurosurg75: 8–14, 1991.
2.
LinTFoxAJDrakeCG: Regrowth of aneurysm sacs from residual neck following aneurysm clipping. J Neurosurg70: 556–560, 1989.
3.
RaymondJRoyD: Safety and efficacy of endovascular treatment of acutely ruptured aneurysm. Neurosurgery41: 1235–1246, 1997.
4.
GurianJHMartinNAKingWA: Neurosurgical management of cerebral aneurysms following unsuccessful or incomplete endovascular embolization. J Neurosurg83: 843–853, 1995.
5.
Fernandez-ZubillagaAF: Endovascular occlusion of intracranial aneurysms with electrically detachable coils: Correlation of aneurysm neck size and treatment results. Am J Neuroradiol15: 815–820, 1994.