Abstract
Introduction
The Flow Re-direction Endoluminal Device (FRED) is a novel flow diverter with a unique double stent design, with an inner stent composed of 48 nitinol wires, and an outer stent with 16 nitinol wires. It is designed for endovascular cerebral aneurysm treatment, although, limited data exist regarding in-stent stenosis (ISS) rates associated with FRED devices.
Methods
A registry encompassing two North American comprehensive stroke centers was the base of this study. We longitudinally assessed patients implanted with FRED devices, emphasizing baseline demographics, aneurysmal characteristics, procedural data, aneurysmal occlusion, and the incidence of ISS.
Results
In our cohort of 87 patients receiving 88 FRED devices, ISS occurred in 10.4% (9) of cases. Mild ISS (<50%) was noted in 8.0% (7) of patients, moderate ISS (50–75%) in 1.1% (1), and severe ISS (>75%) in 1.1% (1). Analysis indicated ISS in 17.0% (8) of patients with the FRED device and 5.7% (2) with the FRED-X device; all ISS cases in the FRED-X group were mild. Differences in ISS rates between device types were not significant (p = 0.122). Delayed thrombotic events were documented in 6.9% (6) of patients. Aneurysm occlusion rates, measured via the Raymond–Roy Scale (RRS), showed adequate occlusion (RRS 1 or 2) in 68.8% at 3 months, 74.6% at 6 months, and 89.3% at 12 months.
Conclusions
The study elucidates the efficacy and safety profile of FRED devices, presenting favorable aneurysm occlusion rates and low retreatment needs while underscoring the manageability of ISS.
Keywords
Get full access to this article
View all access options for this article.
References
Supplementary Material
Please find the following supplemental material available below.
For Open Access articles published under a Creative Commons License, all supplemental material carries the same license as the article it is associated with.
For non-Open Access articles published, all supplemental material carries a non-exclusive license, and permission requests for re-use of supplemental material or any part of supplemental material shall be sent directly to the copyright owner as specified in the copyright notice associated with the article.