A 43-year-old woman attended for stent assisted coiling. A Neuroform 30 × 4.5 mm stent had been successfully placed over the left periophthalmic aneurysm. During the coiling the first coil migrated through the crowns in the stent, lodging at the MCA bifurcation. We believe that the coil herniated through the overlying stent due to the carotid siphon curvature and the open cell design. Furthermore the distal markers of the stent impeded coil extraction with a MERCI device.
MoretJCognardC: Reconstruction technic in the treatment of wide-neck intracranial aneurysms. Long-term angiographic and clinical results. Apropos of 56 cases. J Neuroradiol24: 30–44, 1997 (in French).
2.
WankeIDoerflerA: Treatment of wide-necked intracranial aneurysms with a self-expanding stent system: Initial clinical experience. Am J Neuroradiol24: 1192–1199, 2003.
SaniSJobeKWLopesDK: Treatment of wide-necked cerebral aneurysms with the Neuroform 2 Treo stent. A prospective 6-month study. Neurosurg Focus15: 18(2): E4, 2005.
5.
BroadbentLPMoranCJ: Management of neuroform stent dislodgement and misplacement. Am J Neuroradiol24: 1819–1822, 2003.
6.
HsuSWChaloupkaJC: In vitro studies of the neuroform microstent using transparent human intracranial arteries. Am J Neuroradiol27: 1135–1139, 2006.
7.
EbrahimiNClausB: Stent conformity in curved vascular models with simulated aneurysm necks using flat-panel CT an in vitro study. Am J Neuroradiol28: 823–829, 2007.
8.
BenndorfGClausB: Increased cell opening and prolapse of struts of a neuroform stent in curved vasculature: Value of angiographic computed tomography: Technical case report. Neurosurgery58(4 Sup), 2006.
9.
RichterGEngelhornT: Flat panel detector angiographic CT for stent-assisted coil embolization of broad-based cerebral aneurysms. Am J Neuroradiol28: 1902–1908, 2007.