Abstract
Objective
Visceral ischemia can be a potentially life-threatening complication of intra-aortic balloon pump (IABP) support. A shorter IABP catheter might lead to a reduction of visceral complications. In this animal study, we evaluate the effects of a 35-mL short catheter in comparison with a 40-mL standard-sized catheter.
Methods
Eighteen healthy swine underwent 120-minute ligation of the left anterior descending coronary artery followed by 6 hours of reperfusion being supported by either a short IABP catheter (short group) (n = 6) or a long IABP catheter (long group) (n = 6) or with no assistance (controls) (n = 6). Hemodynamics, visceral and coronary flows, as well as biochemical markers were evaluated throughout the different phases of the protocol.
Results
Mesenteric flows increased significantly at reperfusion (P < 0.001 both) remaining constant afterward (all, P > 0.05) in the short group, while remaining significantly lower in the long group at the start of reperfusion, remaining constantly lower than the short group and controls (P < 0.001 vs short, P < 0.003 vs controls). In both long and short groups, catheters improved renal flows at reperfusion (P < 0.001 both) without any further variation (P < 0.05). In the short group, the flows were higher during the whole of reperfusion (all, P < 0.05). Intra-aortic balloon pump support improved hemodynamic indices and coronary blood flows during reperfusion to a similar extent in both the small and the long group (P > 0.05).
Conclusions
The short IABP catheter proved to be as effective as the standard-sized catheter in supporting hemodynamics and coronary circulation. Furthermore, it even improves visceral flows in comparison with conventional IABP catheters.
Get full access to this article
View all access options for this article.