Abstract
Introduction
Ankle/brachial indices are inaccurate in the presence of calcification, and physicians may rely on the measurement of digital pressures. As the population continues to age and with the escalation in type 2 diabetes, the importance of, and reliance on, toe pressure measurements will increase. The aim of this study was to assess the reproducibility of toe pressure measurements in a single vascular laboratory.
Methods
Repeated ankle/brachial indices and toe pressures were measured in 20 patients (10 with known peripheral vascular disease and 10 with aneurysmal or carotid artery disease but no history of PVD), and 10 control patients. Three measurements were made 48 hours apart. All measurements were made by a single vascular technologist. Reproducibility was assessed by the use of the repeatability coefficient and the intraclass correlation coefficient.
Results
Ankle/brachial indices ranged from 0.36 to 2.4, toe pressures from 18 mmHg to 173 mmHg, and toe/brachial indices from 0.11 to 1.1. The repeatability coefficient showed no significant difference between measurements (p > 0.1) and the intraclass correlation coefficient estimates showed high agreement between repeated measurements (0.77–0.99). Bland-Altman plots indicated that the observer variability was equally distributed across the range of pressure measurements.
Conclusion
These results confirm the intraobserver reproducibility of toe pressure measurements; however, further work is required to demonstrate inter-observer reproducibility.