Abstract
Background
In-stent restenosis remains a common and important complication after endovascular treatment of superficial femoral artery peripheral artery disease. It occurs in 14 to 35% of cases in 1 year and there is still no efficient treatment for this condition. Paclitaxel-coated balloons have shown promising results.
Objective
Investigate the 3 year results of superficial femoral artery in-stent restenosis treated with paclitaxel-coated balloon angioplasty, using the Lutonix™ 035 device.
Methods
We conducted a retrospective observational study with patients with symptomatic (Rutherford 2 to 5) superficial femoral artery in-stent restenosis, that were treated with paclitaxel-coated balloon angioplasty using the Lutonix™ 035 device, in a single center from January 2016 to December 2020. Duplex scan was used to follow the patients. Primary patency was obtained through Kaplan-Meier analysis. Mortality, and amputation rates were also evaluated.
Results
105 patients were included. Two patients had technical failure and required an additional stent, and were thus excluded. 103 patients were analyzed. Primary patency was 91.26, 80.47, and 67.71%, respectively, in the first, second, and third year after the procedure. There were no deaths 30 days after the procedure. There were no major amputations during the 3 year follow-up.
Conclusion
Paclitaxel-coated balloon angioplasty with the Lutonix™ 035 device was a safe and effective treatment to superficial femoral artery in-stent restenoses. The results were maintained along the 3 year follow-up.
Keywords