Abstract
Purpose:
To develop an animal model for the acute and chronic monitoring of pressure within abdominal aortic aneurysms (AAAs) to be treated with endovascular grafts.
Methods:
A strain-gauge pressure transducer was placed within an AAA created from a prosthetic vascular graft. Prosthetic aneurysms were implanted into 17 canine infrarenal aortas. The intra-aneurysmal pressure was monitored and correlated with noninvasive forelimb sphygmomanometry for 2 weeks. After this time, an intravascular manometer catheter was passed into the aneurysm. Simultaneous pressure measurements were obtained using the implanted strain-gauge pressure transducer, the manometer catheter, and the forelimb sphygmomanometer. Angiography was performed to assess intraluminal morphology, aneurysm anastomoses, and adjoining aortic vessels. In addition, two control animals underwent intra-aneurysmal pressure monitoring after standard surgical aneurysm repair.
Results:
There was excellent correlation (r = 0.97) between the pressure measurements obtained with the implanted strain-gauge pressure transducer and the intravascular manometer. Close correlation was also observed between the implanted strain-gauge transducer and the forelimb sphygmomanometer (r = 0.88) during postprocedural monitoring. Intra-aneurysmal pressure was lowered dramatically by surgical exclusion (aneurysm: 15/5 ± 7/4 mmHg; systemic: 124/66 ± 34/17 mmHg; p < 0.001). The prosthetic aneurysms were successfully imaged with angiography.
Conclusions:
This animal model provides an accurate and reproducible means for measuring intra-aneurysmal pressure on an acute and chronic basis. It may be possible to use this model in the assessment of endovascular devices to determine their efficacy in reducing intra-aneurysmal pressure. Evaluation of complications associated with their use, such as patent aneurysm side branches, perigraft channels, and perianastomotic reflux, may also be possible.
Keywords
Get full access to this article
View all access options for this article.
